An index matching liquid is used to suppress reflections when illuminating optical glass with a light source (this is particularly important when the light source is a laser). A nonvolatile liquid with a refractive index close to that of the glass is selected in order to reduce the difference in refractive index between the sample cell and its surroundings. So long as the condition Δn(glass-liquid) << Δn(glass-air) is satisfied, this reduces the visibility of the interface, by minimizing the optical contrast between the glass and the continuum surrounding it. Ideally Δn should be as close to zero as possible.

Brookhaven’s BI-200SM Research Goniometer uses decalin as an index matching liquid, as it is both nonvolatile and has a refractive index close to that of laboratory grade optical glass (n ~1.48). A specialized optical glass vat is used to hold the decalin that the cylindrical sample cell is placed into. A goniometer measures across a continuum of scattering angles, meaning that in contrast to a benchtop light scattering system, a square cuvette cannot be used as the vast majority of scattering angles would be optically inaccessible (aside from of 90 degrees, backscatter, and forward scatter). Since the goniometer uses a cylindrical sample cell, index-matching is required in order to not reflect laser light towards the detector.
The 200SM uses a circulating pump (FC) to filter the decalin in order to remove any dust or debris that otherwise might be transferred into the index-matching vat when sample cells are loaded/unloaded. The condition of the decalin should be checked periodically, replacing in-line filters as needed.
# # #
Brookhaven Instruments are pioneers in the field of light scattering for nanoparticle and biomolecule size determination. For over 40 years we have been innovating new advances in the field. We are trusted experts who listen first to understand our customers’ challenges and then configure appropriate light scattering systems to meet our customers’ needs. We also offer dedicated support resources to help you along the way. We are real people serving real people in realtime.TM Our instruments have always been designed and built in the USA.
Contact us today to learn more about our light scattering solutions for particle size and zeta potential analysis.