Home > Products > BI-MwA – Molecular Weight Analyzer
BI-MwA – Molecular Weight Analyzer
Quickly, easily, and accurately determine the absolute molecular weights of proteins and polymers
The BI-MwA has the highest performance/price ratio of any light scattering detector used for molecular weight determination. Eliminate SEC/GPC column calibration and improve data quality. Rapidly reduce uncertainty with no assumptions required and true extrapolation to zero angle.
Simple to use, sophisticated features
Inject your sample into the low-volume, 7-angle flow cell. The sample is illuminated by a temperature-stabilized, precision power-controlled diode laser. The ultra-stable, high-sensitivity, low-noise CCD detector automatically collects the scattered light. Then, the software extrapolates the data to zero angle for the absolute molecular weight determination.
Features at a glance
Multiangle, laser light scattering detector for SEC/GPC or stand-alone use
Proteins & polymers
Polysaccharides & biopolymers
Study protein aggregation, oligomerization, & complex formation
Determine the molecular weight distribution (flow mode) or determine the average molecular weight (batch mode). Macromolecular samples often have a range of molecular weights. In some cases, this distribution is quite narrow, and in other cases, it is very broad and/or multimodal. Variations in the distribution can indicate the presence of impurities or aggregation. However, for other applications, the average molecular weight is sufficient to characterize your sample. Sometimes the distribution is of interest, and sometimes the overall average is sufficient. Either way, the BI-MwA is the right tool for the job.
Chromatography (GPC/SEC)
Molecular Weight Distribution (Flow Mode)
When the distribution of molecular weights in your sample is important, consider the BI-MwA as a chromatography detector for your SEC/GPC system. In this way, you can readily determine the true molecular weight distribution. Many average values unavailable from a batch mode measurement can be calculated from these distributions. Better yet, the molecular weight distribution is determined while the problems of column calibration are avoided.
The way it was:
Size exclusion chromatography (SEC), also known as gel permeation chromatography (GPC), uses columns to separate polymer samples. In the standard method, columns are calibrated to obtain a relationship between molecular weight and elution volume. However, the column-sample interaction depends on not just the size (molecular weight) but also the chemistry of the sample. Therefore, an accurate column calibration requires standards over a range of molecular weights with the same chemistry and structure (e.g., branching) as the sample. With a few exceptions, such standards are difficult or impossible to obtain, especially for new or unique materials.
A better approach:
The BI-MwA provides a method of determining absolute molecular weights without resorting to any assumptions about the sample or column calibration. Simply attach the BI-MwA and an appropriate concentration detector such as the BI-DNDC and the ParSEC GPC software to the end of the SEC column. The column is then used to separate the species of interest and the BI-MwA gives full information on the molecular weight of each fraction. Figures 1 and 2 show the results of a measurement of the molecular weight distribution of a dextran sample by SEC.
The BI-MwA can be used as part of a GPC system for absolute MW determination
Batch Mode
Average Molecular Weight (Batch Mode)
When the average molecular weight is sufficient to characterize your sample, use the BI-MwA in batch mode to easily determine average properties. The BI-MwA permits routine determination of the weight average molecular weight, Mw, the z-average radius of gyration, Rg, and the second virial coefficient, A2. Readily prepare Zimm, Berry, or Debye plots to calculate these parameters with the optional BI-ZPMwA software.
Using the BI-MwA to determine the average molecular weight
Obtain molecular weights in batch mode by preparing dilute solutions of known concentration. Follow the steps in the software to enter concentration values and inject solutions. The software will automatically evaluate the data. With the click of a mouse, data are analyzed and presented. Data review and analysis are under user control. Results are presented on screen and in printed reports. See one example of batch mode data in Figure 3.
Figure 3 Zimm plot of a polystyrene sample in toluene. The results are in excellent agreement with expectations. The weight average molecular weight, Mw, and radius of gyration. Rg, and second virial coefficient A2, are rapidly and easily obtained.
Samples are unaffected by measurement and can be recovered for future analysis.
User-Friendly Application Software
The application software is easy to learn, convenient to use, and maximizes user productivity.
The BI-MwA can be used with industry-leading, SEC-system software, which allows rapid measurement setup, user-customizable output, and reliable data collection and analysis.
The batch-mode software can generate Zimm, Debye, and Berry plots. Figure 3 is an example of a Zimm plot obtained with the BI-MwA. For complex samples, both linear and higher-order polynomial fits can be performed. All fitting is coupled with sophisticated statistical analysis. Finally, the software incorporates algorithms for taking advantage of the flow system to remove artifacts due to contaminants (dust). Thus, good information can be recovered even from imperfect samples.
Why Seven Angles?
Absolute molecular weight is determined from the scattered intensity at zero angle. Unfortunately, it is impossible to measure scattered light at zero angle since transmitted light will also be measured. Measurement at one or more nonzero angles and extrapolation to zero angles is necessary for molecular weight determination. A single-angle light scattering instrument relies on extrapolation based on one data point; occasionally, it works. In addition, measurement of the change in scattering with angle is necessary to determine the radius of gyration. Therefore, a multi-angle instrument is far better than a single-angle instrument.
When comparing multi-angle instruments, keep in mind that the precision of the light scattering results is approximately proportional to the square root of the number of angles. Thus, results from a seven-angle instrument such as a BI-MwA are significantly more precise than results from a two- or three-angle instrument. The seven-angle design offers the best performance/price ratio in the industry.
Optimal Design, Small Footprint, Rugged Construction
The BI-MwA design eliminates the pitfalls of similar instruments. To avoid artifacts and drift due to trapped bubbles, the flow path is vertical, not horizontal. The patented cell is more easily flushed clean when samples are changed.
The sample cell can withstand pressures up to 3.5 MPa, a much higher pressure than other instruments of this type. Therefore, the BI-MwA is a perfect complement to an SEC system where high pressures often lead to premature instrument failure.
In addition to the cell design, the overall mechanical design is also optimized for maximum performance. The footprint (21 cm wide x 38 cm deep) conserves valuable bench space. In addition, the internal partition between the electronics and the liquids minimizes the extent of damage and the cost of repair if a leak occurs.
Contact us for complete assistance in setting up a customized SEC system.
Integrated, dedicated, powerful microcontroller. USB driven.
Detector
CCD, ultra-high sensitivity, and spatial uniformity.
Analog Inputs
4 standard, with 16 and 24-bit resolution, and 15 additional analog channels (optional) with 24-bit resolution. Computer selectable gain adjustment. Suitable for use with most common RI, UV, viscometer, thermocouple, thermistor, injector, pressure gauge, and pump outputs.
Power Requirements
100/115/220/240 VAC, 50/60 Hz,25 Watts
Dimensions
Size: 195(H) x 210(W) x 380(D) mm; Weight: 5.5 kg
Certificates
CE Certified
U.S. Patent # 6,052,184
A policy of continued improvement may lead to specification changes.
Speak To An Expert Today About Our Analytical Services
We’re standing by to help you select the right services for your characterization needs.
This website uses cookies to give you the very best experience. By continuing to browse this website you are agreeing to our use of cookies. To find out more how we use cookies, please review our Privacy Notice Accept and Close
Privacy & Cookies Policy
Privacy Overview
This website uses cookies to improve your experience while you navigate through the website. Out of these cookies, the cookies that are categorized as necessary are stored on your browser as they are as essential for the working of basic functionalities of the website. We also use third-party cookies that help us analyze and understand how you use this website. These cookies will be stored in your browser only with your consent. You also have the option to opt-out of these cookies. But opting out of some of these cookies may have an effect on your browsing experience.
Necessary cookies are absolutely essential for the website to function properly. This category only includes cookies that ensures basic functionalities and security features of the website. These cookies do not store any personal information.