BI-870 Dielectric Constant Meter
Measurement of dielectric constants has never been so easy with the BI-870
Simply insert the probe in the liquid to be measured, adjust the two controls on the front panel, and read the dielectric constant from the display. The BI-870 can accurately measure in low and high dielectric solvents, including mixed liquids and solutions.
The BI-870 has two selectable sensitivity ranges: 1-20 and 1-200. Absolute accuracy is ± 2%, while repeatability and linearity accuracy are better than 0.2%. The measurement signal applied to the outer cylinder of the probe is a low-distortion sine wave at a frequency of 10 kHz. The amplitude is approximately 7 volts rms in the 1-20 range and 0.7 volts in the 1-200 range. The frequency is crystal-controlled and is, therefore, stable to approximately 1 part in 105. The dielectric constant of the liquid sample is determined by measuring the current between the outer and inner cylinders of the probe. With a stable voltage source and precisely known probe parameters, it is possible to display the dielectric constant directly. Calibration is simple using the back panel adjustment with a liquid of known dielectric constant.
The probe has an open structure and is easy to clean. It is constructed from two precision cylinders, machined from type 316 stainless steel. Six Teflon balls maintain the cylinder spacing. If the probe is used primarily with low-dielectric constant hydrocarbon fluids, we recommend cleaning by agitation in acetone or ethyl alcohol, followed by gentle drying with clean compressed air. The probe must be cleaned before any residue dries on the cylinders. Since the probe is made of stainless steel and Teflon®, it may be cleaned in almost any solvent.
Good zeta potential determinations of suspended colloidal particles depend on an accurate knowledge of the dielectric constant of the liquid. For a range of particles of interest, several different liquids may be used, including mixtures of various proportions creating the need for an accurate measurement of dielectric constant. While literature values of dielectric constants are suitable when using pure liquids values are not readily available for mixtures of liquids. There are no generally acceptable methods of calculating values for mixtures from pure component values. For example, weighting by mole fraction leads to errors of 17% and 6% for the two mixtures given in the table below. Weighting by volume leads to errors of -8% for Methanol/Water and 2.4% for Dichloro Methane/Benzene respectively.
Measurement Principles
Operation
The BI-870 has two selectable sensitivity ranges: 1-20 and 1-200. Absolute accuracy is ± 2% while repeatability and linearity are better than 0.2%. The measurement signal applied to the outer cylinder of the probe is a low-distortion sine wave at a frequency of 10 kHz. The amplitude is approximately 7 volts rms in the 1-20 range and 0.7 volts in the 1-200 range. The frequency is crystal-controlled and is, therefore, stable to approximately 1 part in 105. The dielectric constant of the liquid sample is determined by measuring the current between the outer and inner cylinders of the probe. With a stable voltage source and precisely known probe parameters, it is possible to display the dielectric constant directly. Calibration is simple using the back panel adjustment with a liquid of known dielectric constant.
Probe
The probe has an open structure and is easy to clean. It is constructed from two precision cylinders, machined from type 316 stainless steel. Six Teflon balls maintain the cylinder spacing. If the probe is used primarily with low-dielectric constant hydrocarbon fluids, we recommend cleaning by agitation in acetone or ethyl alcohol, followed by gentle drying with clean compressed air. It is important that the probe be cleaned before any residue dries on the cylinders. Since the probe is made of stainless steel and Teflon®, it may be cleaned in almost any solvent.
BI-870 Dielectric Constant Measurements for Common Liquids
Material
Temperature
Measured Dielectric Constant
Published1 Dielectric Constant
% Difference2
Cyclohexane
250C
2.03
2.015
0.7
Ethyl acetate
250C
6.05
6.02
0.5
Dichloro methane
250C
8.92
9.073
-1.6
Methanol
250C
33.2
32.63
1.7
De-ionized Water
250C 240C 230C
79.3 79.9 80.2
78.54 78.90 79.27
1.0 1.3 1.2
Dichloro methane + Cyclohexane4
250C
5.35
N/A
–
Methanol + Water4
250C
61.2
N/A
–
1. CRC Handbook of Chemistry and Physics. 2. %Difference = [(Measured – Published)/Published] x 100. 3. Corrected for temperature to match 25 °C measurements. 4. Liquids were mixed 1:1 by volume at 25 °C.
Specifications
Specifications
Full Scale Sensitivity
Dielectric Constant of 1-20 and 1-200
Maximum Conductivity of Sample
Range 1-20: 1 µS/cm; Range 1-200: 10 µS/cm,(
Accuracy
2% Absolute; Repeatability and Linearity 0.2%
Probe Materials
Stainless Steel and Teflon®
Minimum Liquid Required
~23 mL, standard probe.
Power Requirements
100/115/220/240 VAC, 50/60 Hz, 10 watts
Dimensions
190mm x 240mm x 70mm
Operating Temperature
22-58 °C
Measurement Signal
Low-distortion 10 kHz sine wave
Display
Backlit LCD
Output
Analog Recorder (optional or customer supplied), Full Scale Reading 1.999 V
Calibration
Back Panel Adjustment with Reference Liquid
A policy of continual improvement may lead to specification changes
Speak To An Expert Today About Our Analytical Services
We’re standing by to help you select the right services for your characterization needs.
This website uses cookies to give you the very best experience. By continuing to browse this website you are agreeing to our use of cookies. To find out more how we use cookies, please review our Privacy Notice Accept and Close
Privacy & Cookies Policy
Privacy Overview
This website uses cookies to improve your experience while you navigate through the website. Out of these cookies, the cookies that are categorized as necessary are stored on your browser as they are as essential for the working of basic functionalities of the website. We also use third-party cookies that help us analyze and understand how you use this website. These cookies will be stored in your browser only with your consent. You also have the option to opt-out of these cookies. But opting out of some of these cookies may have an effect on your browsing experience.
Necessary cookies are absolutely essential for the website to function properly. This category only includes cookies that ensures basic functionalities and security features of the website. These cookies do not store any personal information.