BI-RIVS

The BI-RIVS is a combination RI and Viscometry detector used in SEC/GPC determinations of the MWD using universal calibration.

Description

Refractometer & Viscometer

The BI-RIVS is a combination RI and Viscometry detector used in SEC/GPC determinations of the MWD using universal calibration.

A true Dual-Detector, refractometer and viscometer, the only one of its kind available. The refractometer is an integral component of the viscometer bridge, measurement of concentration AND viscosity take place on the exact same sample segment, at the very same time. Compared to the more common “external” connection of the two detectors, the BI-RIVS Dual-Detector provides a factor of 2 to 5 higher sensitivity, near ZERO delay volume and unsurpassed overall performance.

There is no need to wait for the sample “breakthrough” (the negative signal you can expect after each injection) or to adapt an “internal delay” to the number of columns you use like other viscometer designs require. The Dual-Detector will match any application you might think of, with any solvent, at any temperature up to 80°C.

Features at a Glance:

  • Dual Detector
  • Zero Dead Volume
  • Highest Sensitivity
  • Up to 80 °C

Specifications

Viscometer Specifications

Differential Pressure Range ±5.0 kPa
Differential Pressure Gains 1, 2, 4, 8
Inlet Pressure Range 0 to 150 kPa
Sample Cell Volume 15 µl
Shear Rate (1.0 ml/min) ≤ 2,700 sec-1
Noise Level 0.2 Pa, diff. Pressure Channel, 25 °C
Analog Output (integrator) 1.0 Volt / 5 kPa FSD Differential Pressure
1.0 Volt / 100 kPa FSD Inlet Pressure


Refractive Index Detector Specifications

Refractive Index Range 1.00 to 1.75
Cell Volume 8 µl
Measurement Ranges (Δ Full Scale) 6×10-4, 3×10-4, 1.5×10-4, 7.5×10-5
Noise Level < 2.5×10-9 ΔRIU, 25 °C
Analog Output (integrator) 1.0 Volt FSD


Common Specifications

Total Detector Volume ≈ 40 ml (including reservoirs)
Temperature Accuracy ± 0.5 °C
Temperature Stability > 0.01 °C
Digital Interface RS-232C
USB
LAN
Data Transfer Rate (baud) 4800, 9600, 19200, 57600, 115200
Digital Inputs Flush, Zero, Sync, Error
Digital Outputs Sync, Error