Executive summary
Measuring size distributions that span nanometers to microns is a unique challenge. Previously, Brookhaven Instruments offered the BI-XDC, a centrifugal instrument that relied on X-Rays to detect sedimenting particles. Here, we discuss several commercially available particle size analyzers, and compare them to the XDC. In comparing their strengths and weaknesses, we find that a Disc Centrifuge Photosedimentometer, such as the BI-DCP, would be an ideal choice to replace the BI-XDC.
Introduction
One of the most difficult decisions for companies and researchers is what to do with a failing piece of equipment when replacements are no longer being manufactured. Many BI-XDC particle size analyzers are nearing the end of their life, requiring the search for a new instrument. This is made more complicated by the fact that a BI-XDC was capable of measuring particles from tens of nanometers in diameter to tens of microns in diameter or more—a technically astounding range—and possessed unique response to metals and metal oxides. Here we provide an overview of some of the commercial, off-the-shelf particle sizing techniques, and investigate their strengths and weaknesses as opposed to a BIXDC instrument.
Industry Overview
There are several particle sizing methods that may be of interest to someone attempting to replace a BI-XDC instrument. Here, we focus on a limited set of off-the-shelf techniques that owners of a BI-XDC may be considering as a replacement; namely, Disc Centrifuge Photosedimentation (DCP), Dynamic Image Analysis (DIA), Laser diffraction, and Dynamic Light Scattering (DLS). Note that these techniques do not require regular calibration, as all but Dynamic Image Analysis are firstprinciples techniques, and Dynamic Image Analysis only requires the size per pixel to be measured once. A summary of these techniques and their strengths and weaknesses is shown in Figure 1:

Name | Advantages | Disadvantages |
|---|---|---|
DCP |
|
|
DIA |
|
|
Laser Diffraction |
|
|
DLS |
|
|
Figure 1: A graphical representation of the size ranges, advantages, and disadvantages to the techniques discussed in this work
Disc Centrifuge Photosedimentometer

The Disc Centrifuge Photosedimentometer (DCP) is a technique that involves spinning a fluid, such as water or another “spin medium”, at a high speed, and then injecting particles into the fluid to analyze into them. A representation is shown in Figure 2. A photogate is placed to view the spin fluid and return the transmission of the fluid. As the particles are subjected to centrifugal and drag forces, they quickly reach a terminal velocity in the fluid. [1] [2] They then pass through the fluid and eventually reach the photogate, where they scatter light from the photogate, causing a drop in transmission. The time from injection until this drop in transmission depends on the mass and volume of the particles. This measured time, along with the density of the particles, is used to calculate the particle size. DCP techniques are suitable for a size range of 10 nm to over 50 µm. [3] [4] [2] One product capable of making these measurements is the BI-DCP.
DCP technology is a robust method of particle analysis. It is agnostic of focus, multiple scattering, and most other optical concerns, making it excellent for highly concentrated samples. It is able to easily measure multimodal and complicated distributions of particle sizes. Finally, its wide range of sizes able to be analyzed (10 nm to over 50 µm) is unmatched by similar analyses. [3]
However, DCP does have some flaws. Data acquisition can take up to 30 minutes to fully characterize a sample. A DCP operator must know the density of their sample for accurate results. The DCP is unable to be used during a production process, so a batch of product must be finished, and then a representative sample must be tested. Finally, the spin fluid must be changed out after taking several measurements, a process that takes a few minutes as well. [3] [4]
Laser Diffraction

Laser diffraction is a technique based on how light scatters off of particles. To make a measurement, a laser beam passes through a stream of particles, and the laser light diffracts off the surface of those particles. A representation of this is shown in Figure 3. The pattern it makes is due to scattering according to the size of the particles. This scattered light is measured, and then fitted to a diffraction pattern. Using theoretical calculations from well-known methods like Fraunhofer or Mie theory, the diffraction pattern is used to back out a spherical diameter. This technique is suitable for generally tens of nanometers to tens of microns. [5]
This comes with several obvious advantages. Data collection is fast and easy—taking a measurement takes around half a second. [6] Furthermore, since this depends on a stream of particles, laser diffraction lends itself to an instrument that can be used in real time with a production process. And, as long as the particles are mostly spherical, Laser diffraction can measure a particle size. [7]
However, Laser diffraction is not always the best option for measuring samples. For one, if the particle sizes vary widely, or if there is a multimodal size distribution, then fitting particle scattering to a particle size may introduce errors. [8] Furthermore, the more aspherical the particles, the more errors will be introduced, since the scattering will not be represented by Mie calculations. [6] [9] And finally, some laser diffraction equipment overstates its ability to analyze nanoscale particles. [10] In summary, careful error management and analysis is necessary when using laser diffraction.
Dynamic Image Analysis

Dynamic Image Analysis (DIA) involves high speed microscopic imagery of particles flowing past an objective. These particles could be aerosolized, or suspended in fluid, and then allowed to flow past a microscope objective, which takes high resolution images of the particles. A representation of this system is shown in Figure 4. The images are quickly analyzed using advanced graphical processes, and the volume, speed, and various shape parameters are measured. This is typically suitable for a size range of 10 to 1,500 microns. [11] [12] [13]
Dynamic Image Analysis has significant advantages. It is able to measure both volume and speed, which would be helpful in determining the density of the particle. Moreover, the multiple shape parameters help in ensuring the particles are correctly formed in shape. Additionally, multimodal distributions are easily measured. Using the computationally intense methods of DIA allows for extensive statistical measures that other techniques may not provide without additional microscopy. Finally, this technique inherently involves a flow cell, which provides the possibility of integrating DIA into a production line. [13]
However, the technique comes with very significant drawbacks. The speed of the medium needs to be carefully controlled, otherwise, the particles may be photographed with motion blur, distorting the measurements. Similarly, the particles must remain in focus. [14] As such, accuracy and error management is imperative in working with DIA. Lastly, it is only available for relatively larger particles with significant accuracy. [15]
Dynamic Light Scattering

Dynamic Light Scattering (DLS) is another technique based on particles scattering light, and has been extensively studied. Rather than expecting Mie scattering, where the particles must be at least several tens of nanometers, DLS is based on Rayleigh scattering, which typically occurs with smaller particles. [16] A DLS instrument passes a laser through a cuvette filled with particles, and the laser light scatters off of the particles. A detector, placed some distance away, measures the intensity of the scattered light over time. As particles enter and exit the laser beam under the influence of Brownian motion, they cause the scattered light to increase or decrease, resulting in a fluctuating intensity. [16] These intensity fluctuations are used to construct an autocorrelation function, which is then used to determine the speed of the Brownian diffusion, which is then used to calculate hydrodynamic size. A representation of this process is given in Figure 5 This measurement is suitable for particles of less than 10 nm to 10 µm, with decreasing accuracy as the particles grow larger than 2 microns. [16] [17]
DLS has multiple unique benefits. It is uniquely suited to measuring nanometer-scale particles. [18] [19] It is also relatively fast, with typical measurements taking under 10 minutes. Furthermore, applying an electric field using a minimal amount of additional hardware allows for zeta potential measurement, a measurement of the charge on the particles. [20] This makes it ideal for fully characterizing very small particles.
However, DLS does have some drawbacks. It struggles with heterogenous samples and multimodal distributions, with a general guideline of 2-3 peaks in the population at most. [21] It also has significant difficulties with larger particles—as the particles grow in size to greater than about 2 microns, DLS is unable to sufficiently fit the scattering fluctuations with the autocorrelation function. [17] Thus, while the DLS technique is powerful for small particles, it has limitations.
BI-XDC Compared to other techniques
XDC was a form of centrifugal particle size characterization, where a particle solution was injected into a centrifuge, much like DCP. XDC then used an X-Ray beam to measure the transmission coefficient of the spin fluid, and used this to determine the particle size, the same as a DCP. This is depicted in Figure 6. X-rays were especially sensitive to highly conductive particles, compared to water, which was transparent for lower energy X-rays. This allowed metal and metal oxide particles to be well characterized by XDC.

Comparing the XDC to all of the methods previously considered, then, a replacement would typically need to be able to characterize a large array of sizes, allow for the characterization of highly conductive particles, and be agnostic to concerns about multiple scattering and other complicated optical behavior. The most similar technology to an XDC is the DCP. A DCP system is different in its sensing methods in that light scatters from more materials than metals and metal oxides. However, metal oxides are still highly absorbing in the optical range, and water is transparent, making measurements experimentally similar.
Since the principles of operation for the DCP are substantially similar to that of the BI-XDC, a BI-XDC operator would be very confident in moving to DCP product from a XDC instrument. Processes and operation of the instrument would already be well suited for a DCP. For example, the comparative graph shown in Figure 7 depicts data obtained in previous measurements with a XDC measurements [4]. The process was repeated for the DCP with a similar, but not the same, ZnO sample. Namely, ZnO was prepared at concentrations of 2% (w/v) in 0.1% (w/v) tetrasodium pyrophosphate (TSPP). DCP data for the ZnO was taken, and is displayed with historic XDC data in Figure 7. The data shows a high degree of similarity, with the DCP measurement having a similar size distribution, shifted slightly larger. This is likely due to different sample sources. We note that in both cases, the extended nature of the ZnO particle size distribution was fully captured by the instrument.

Conclusion
When choosing an instrument to replace the XDC, there are multiple options, each with a particular set of uses. For particles with sizes between tens of nanometers to tens of microns, especially with complicated size distributions, the DCP method is a powerful tool with few competitors. Furthermore, the BI-DCP is a direct adaptation of the same technology used in the XDC, making it easy to incorporate in existing processes. Therefore, the BI-DCP has unique advantages that make it particularly useful for replacing an XDC unit.
Works Cited
[1] C. Ambler, “Theory of Centrifugation,” Industrial & Chemical Engineering, 1961.
[2] B. B. Weiner, “Chapter VII: Sedimentation & Centrifugation,” in Let There Be Light: Characterizing Physical Properties of Colloids, Nanoparticles & Proteins Using Light Scattering, 2019.
[3] M. Kamiti, D. Boldridge, L. M. Ndoping and E. E. Remsen, “Simultaneous Absolute Determination of Particle Size and Effective Density of Submicron Colloids by Disc Centrifuge Photosedimentometry,” Analytical Chemistry, vol. 84, pp. 10526-10530, 2012.
[4] Brookhaven Instrument Corporation, “Comparison of Technologies for Particle Sizing: Dynamic Light Scattering (DLS) vs Centrifugation,” 19 February 2025. [Online]. https://www.brookhaveninstruments.com/comparison-of-technologies-for-particle-sizing-dynamic-lightscattering-dls-vs-disc-centrifugation/. Available:
[5] H. Li, J. Li, J. Bodycomb and G. S. Patience, “Experimental Methods in Chemical Engineering: Particle Size Distribution by Laser Diffraction—PSD,” The Canadian Journal of Chemical Engineering, vol. 97, no. 7, pp. 1974-1981, 2019.
[6] P. Kippax, “Appraisal of the Laser Diffraction Particle-Sizing Technique,” Pharmaceutical Technology, pp. 88-96, 2005.
[7] Z. Ma, H. G. Merkus, J. G. de Smet, C. Heffels and B. Scarlett, “New developments in particle characterization by laser diffraction: size and shape,” Powder Technology, vol. 111, no. 1-2, pp. 55-78, 2000.
[8] A. R. Bah, “Fitting Performance of Particle-size Distribution Models on Data Derived by Conventional and Laser Diffraction Techniques,” Soil Science Society of America Journal , vol. 73, no. 4, pp. 1101-1107, 2009.
[9] R. N. Nelly and J. Kazanjian, “Commerical reference shape standards use in the in the study of particle shape effect on laser diffraction particle size analysis,” AAPS PharmSciTech, vol. 7, 2006.
[10] C. M. Keck and R. H. Mueller, “Size analysis of submicron particles by laser diffractometry – 90% of the published measurements are false,” International Journal of Pharmaceutics, vol. 355, no. 1-2, pp. 150-163, 2008.
[11] D. Tysmans, P. Claeys, L. Deriemaeker, D. Maes, R. Finsy and M. Van Molle, “Size and shape analysis of Sedimentary Grains by Automated Dynamic Image Analysis,” Particle & Particle Systems Characterization, no. 23, pp. 381-387, 2007.
[12] Y. Sun, Z. Cai and J. Fu, “Particle morphomics by high-throughput dynamic image analysis,” Scientific reports, vol. 9, p. 9591, 2019.
[13] M.-N. Pons and J. Dodds, “Chapter Fifteen – Particle Shape Characterization by Image Analysis,” in Progress in Filtration and Separation, Academic Press, 2015, pp. 609-636.
[14] J. List, U. Koehler and W. Witt, “Dynamic Image Analysis extended to Fine and Coarse Particles,” Particulate Systems Analysis, pp. 1-5, 2011.
[15] J. G. Whiting, V. N. Tondare, J. H. J. Scott, T. Q. Phan and M. A. Donmez, “Uncertainty of particle size measurements using dynamic image analysis,” CIRP Annals, vol. 68, no. 1, pp. 531-534, 2019.
[16] P. A. Hassan, S. Rana and G. Verma, “Making Sense of Brownian Motion: Colloid Characterization by Dynamic Light Scatering,” Langmuir, vol. 31, pp. 3-12, 2015.
[17] Z. Jian, J. Li, L. Gao, D. Yang and A. Kananev, “Dynamic Light Scattering: A Powerful Tool for In Situ Nanoparticle Sizing,” Colloids and interfaces, vol. 7, no. 1, p. 15, 2023.
[18] T. G. F. Souza, V. S. T. Ciminelli and N. D. S. Mohallem, “A comparison of TEM and DLS methods to characterize size distribution of ceramic nanoparticles,” Journal of Physics: Conference Series, vol. 733, no. 1, p. 012039, 2016.
[19] C. M. Hoo, N. Starostin, P. West and M. L. Mecartney, “A comparison of atomic force microscopy (AFM and dynamic light scattering (DLS) methods to characterize nanoparticle size distributions,” Journal of Nanoparticle Research, vol. 10, pp. 89-96, 2008.
[20] V. Uskokovic, “Dynamic Light Scattering Based Microelectrophoresis: Main Prospects and Limitations,” Journal of Dispersion Science and Technology, vol. 33, no. 12, pp. 1762-1786, 2011.
[21] F. Varenne, A. Makky, M. Gaucher-Delmas, F. Violleau and C. Vauthier, “Multimodal Dispersion of Nanoparticles: A Comprehensive Evaluation of Size Distribution with 9 Size Measurement Methods,” Springer Nature Link, vol. 33, pp. 1220-1234, 2016.
[22] B. I. Corporation, “The X-Ray Disc Centrifuge – A Unique Particle Sizing Technique,” 29 January 2020. [Online]. Available: https://www.brookhaveninstruments.com/the-x-ray-disc-centrifuge-a-unique-particle-sizing-technique/.