Measurement of Cannabinoid Nanoemulsion Stability with Zeta Potential

NanoBrook is part of a family of benchtop zeta potential analyzers, many of which also include sizing capability. By measuring the zeta potential of droplets in micro- and nano-emulsions, it is possible to predict the shelf stability of a given formulation.

There are a number of models available in the NanoBrook family, the most popular ones  for Zeta Potential include the ZetaPALS, 90Plus PALS and the NanoBrook Omni, all of which include Phase Analysis Light Scattering (PALS), an ultra-sensitive method for measuring zeta potential.

The 90Plus PALS and Omni  have the ability to measure particle size by Dynamic Light Scattering (DLS).

The NanoBrook is an open platform that allows the instrument to be upgraded at any time.

 

  • True PALS for high sensitivity electrophoretic light scattering.
  • High power 40 mW laser light source
  • Dedicated reference beam path for ELS/PALS
  • Palladium Electrode

Zeta Potential (ζ) for Predicting the Stability of Emulsions

A cannabinoid nanoemulsion is a dispersion of oil droplets suspended in aqueous solution. Oil droplets are generally metastable, meaning that in the absence of any sort of stabilizing force will tend to coalesce into larger and larger droplets, eventually leading to visible phase separation.

For the extract to be shelf stable it must be able to resist these attractive surface forces that would otherwise drive droplets towards one another.

The most effective method to impart colloidal stability onto a dispersion of oil droplets is via addition of an amphiphilic molecule which can encapsulate the oil droplet, allowing it to be dispersed in an aqueous (polar)  medium.

These systems are commonly used in food, pharmaceutical, and CBD encapsulation.1 Commonly used encapsulating agents include liposomes, as well as a number of commercial nonionic triblock copolymers.

The latter form polymer micelles, which sequester their hydrophobic blocks on the interior, avoiding exposure to the aqueous medium, and can be used to solubilize sparingly soluble small molecule cargos such as cannabinoids.

These commercially available triblock copolymers are frequently chemically modified to contain terminal ionic headgroups further stabilizing the polymer micelles against aggregation,2 whereas lipid-based encapsulating agents tend to former self-assemble into vesicle like structures.

Amphiphilic molecules can self-assemble into a wide variety of different structures each of which would result in a slightly different encapsulation mechanism. Illustration reproduced from Deng, Foods 10.6 (2021): 1354.

While most surfactants are capable of encapsulating nonpolar cargos including charged and nonionic surfactants, the most effective method to prevent droplet coalescence is via electrostatic stabilization (repulsion of surfaces with like-charges). Typically, this requires a formal surface charge to be imparted to the surface of the droplet.  This surface charge will create  a repulsive force that will limit the tendency of two droplets to approach each other, and thus preventing  physical contact that leads to coalesce.  Careful selection of an emulsifying agent will limit the growth of droplets, resulting in a surfactant stabilized emulsion with droplet dimensions from  tens to hundreds of nanometers.

Zeta potentials of less than 10 mV provide a minimal barrier to aggregation and droplet coalescence. Higher zeta potentials serve as the basis for electrostatic stabilization.

When the mixture cannabinoid extract and carrier oil are encapsulated using charge carrying surfactants, these nanoemulsions will move in response to an electric field. Such systems are easily characterized by electrophoretic light scattering, either by ELS or PALS. The magnitude of the measured zeta potential obtained from light scattering is generally predictive of the stability of a given emulsion.  The zeta potential (ζ) threshold for electrostatic stabilization is greater than 10-20 mV.

A nanoemulsion with a high conductivity presents challenges for traditional ELS measurements. The PALS technique is preferred for samples with high conductivity since it is more salt tolerant and provides for a higher sensitivity.

 

How is Zeta Potential / Surface Charge Measured?

There are two common methods for measuring zeta potential from light scattering, both of which are based on electrophoretic light scattering.  With both methods a sample is added to a cuvette and then blackened palladium electrode is inserted. A laser light source is used to measure the light scattered by particles placed in an electric field. The first method is known as laser doppler electrophoresis which uses a DC voltage to influence the movement of the cannabinoid droplets in the nanoemulsion.  The electrophoretic velocity, or velocity of the particle in an electric field, is then normalized to field strength giving an electrophoretic mobility, and then finally used to calculate a zeta potential from one of several standard models (Smoluchowski, Hückel, or Henry).3

This method, typically referred to as ELS, works well for low conductivity samples, including standard colloids and nanoparticles.  The second method, Phase-Analysis Light Scattering (PALS),4 is based on similar principles, but instead utilizes an AC  voltage. The use of AC allows for a more sensitive measurement to reliably measure much lower magnitude zeta potentials. The PALS method is suitable for samples that have much higher conductivities including those that are at physiological ionic strength or higher.

The most commonly used of the three models to calculate zeta potential is the Smoluchowski equation, where the zeta potential, ζ, can be calculated from knowledge of the electrophoretic mobility, μelec, as well as the viscosity, η, and the dielectric of the continuum liquid, εo as follows:

z = melec* h/ eo

All three of models, Zeta PALS, 90Plus PALS, and Omni, have both ELS and PALS measurement modes with a preference based on sample conductivity.

Light Scattering Optics

In light scattering instruments there are multiple angles used for detection. In Dynamic Light Scattering (DLS) the measurement of cannabinoid droplet size can benefit from the availability of several different scattering angles (θ = 15o, 90o, and 173o ). Zeta Potential Analyzers on the other hand, typically only make use of forward scatter angles (θ < 15o).  This is done to minimize the effects of diffusion broadening.  All of Brookhaven’s NanoBrook series Zeta Potential Analyzers utilize a dedicated reference beam path, where the reference frequency, or demodulation frequency in the case of PALS, is set via a piezoelectric modulator.

All NanoBrook Zeta Potential Analyzers use a dedicated reference beam path for both ELS & PALS.  Both measurements are commonly performed using an Uzgiris electrode.

The reference and sample scattering signals are then combined prior to being detected by the 15 o fiber. The detector is a low dead-time avalanche photodiode detector (APD). The APD module is an advanced single-photon counting device, that replaces photomultiplier tubes (PMT) as the photon counting devices of choice for light scattering instruments.  Modern APD’s have lower deadtimes (< 25 ns) and have greater sensitivity than a PMT.

DLS Specifications - NanoBrook 90Plus PALS, and Omni

DLS Specifications – NanoBrook 90Plus PALS, and Omni
Size Range*NanoBrook 90Plus PALS: 1 nm to 6 μm diameter

NanoBrook Omni: <0.3 nm to 10 μm diameter

 

Concentration Range*NanoBrook 90Plus PALS: 2 ppm to 50 mg/mL

NanoBrook Omni: 0.1 ppm to 50 mg/mL

 

DLS Precision± 1% typically
DLS TechniqueDynamic Light Scattering (DLS)

 

 

Detection AnglesNanoBrook 90Plus PALS: 15° & 90°

NanoBrook Omni: 15°, 90°, & 173°

 

Algorithms and ModelsNNLS, Contin, Cumulants, Lognormal
DLS Standards Conformity ISO13321 and ISO2241
 

Zeta Potential Specifications – NanoBrook ZetaPALS, 90Plus PALS and Omni

Suitable Range*1 nm to 100 μm
Mobility Range109 to 107 m2 /V•s ELS; 1011 to 107 m2 /V•s PALS
Zeta Potential Range*-500 mV to 500 mV
Maximum Sample Conductivity*220 mS/cm
Detection Angle15°
Algorithms and ModelsLaser Doppler Electrophoresis (ELS), Phase Analysis Light Scattering (PALS),

electrophoretic mobility, zeta potential using

Smoluchowski, Hückel, or Henry

Zeta Potential ElectrodeBI-SREL Solvent Resistant Electrode, PEEK housing with Palladium electrodes (included).
 

General Specifications – NanoBrook ZetaPALS, 90Plus PALS, and Omni

LaserStandard Laser 40 mW red diode laser, nominal

640 nm wavelength

Condensation ControlPurge facility using dry air or nitrogen
Temperature Control-5 °C to 110 °C, active control. No external

circulator required.

CorrelatorBrookhaven’s TurboCorr, multi-τ, research grade with 510 hardware channels, 100% efficiency, real-time operation over the entire delay-time range.
Sample CellsDLS: 1 to 3 mL disposable plastic, 50 μL  disposable, 40 μL quartz flow cell, 10 μL quartz minimum.

Zeta Potential: 180 μL, 600 μL, 1250 μL

 

SoftwareWindows compatible Particle Solutions software (included)
Power Requirements100/115/220/240 VAC, 50/60 Hz, 150 Watts
Dimensions23.3 x 42.7 x 48.1 cm (HWD)
Weight15 kg
Environmental

Characteristics

Temperature 10 °C to 75 °C

Humidity 0% to 95%, non-condensing

CE CertificateClass I laser product, EN 60825-1:2001, CDRH
ArchitectureOpen platform that can be upgraded at any time.
AccessoryBI-ZTU Autotitrator for zeta potential or size measurements as a function of pH or additive concentration (optional).
Notes* sample dependent
 

Ordering Information:

The NanoBrook ZetaPALS, 90PlusPALS, and Omni bundles include (3) BI-SCP boxes of 100 disposable plastic cuvettes, BI-SVK92 particle size validation kit, power cord, Particle Solutions software, and instruction manual.

 

All NanoBrook Zeta Potential Analyzers include the BI-SREL zeta potential electrode, BI-ELECCK electrode cleaning kit, (1) BI-SCGO box of 10 glass cuvettes for use with BI-SREL, and BI-ZR5 zeta potential validation kit.

 

References

  1. Deng, Lingli. “Current progress in the utilization of soy-based emulsifiers in food applications—A Review.” Foods10.6 (2021): 1354.
  2. 2. Liu, Yuling, et al. “Redox-sensitive Pluronic F127-tocopherol micelles: synthesis, characterization, and cytotoxicity evaluation.” International journal of nanomedicine12 (2017): 2635.
  3. Uzgiris, E. E. “Laser doppler spectrometer for study of electrokinetic phenomena.” Review of Scientific Instruments 45.1 (1974): 74-80.
  4. McNeil-Watson, Fraser, Walther Tscharnuter, and John Miller. “A new instrument for the measurement of very small electrophoretic mobilities using phase analysis light scattering (PALS).” Colloids and Surfaces A: Physicochemical and Engineering Aspects140.1-3 (1998): 53-57.

 

See Also: